|
As reported, the surgeon did not cut the inflation tube at skin level below the yellow anchor prior to removal as prescribed in the instructions-for-use.Pulling the inflation tube through the abdomen with the yellow anchor attached can lead to the yellow anchor getting caught in the anatomy.Had the user followed this instruction the yellow anchor portion of the inflation tube would have been removed prior to pulling it through the abdomen.Based on the information provided the reported event is determined to be use related as the instructions-for-use were not followed.To date there have been no other reported complaints for this lot number.Per the instructions-for-use, supplied with the device, ¿to deflate the echo ps positioning system release the clamp on the inflation tube, cut the tube as close to the skin as possible, and then discard the excess tube¿.Device not returned.
|