| Model Number |
53618 |
| Medical Device Problem Code |
Sharp Edges (4013)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
12/15/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that male intermittent catheter cut the patient and made bleed.No medical intervention was reported.Per follow up via phone on 22-dec-2022, it was stated that the issue had been resolved and the bleeding resolved on its own.
|
| |
|
Event or Problem Description
|
|
It was reported that male intermittent catheter cut the patient and made bleed.No medical intervention was reported.Per follow up via phone on 22dec2022, it was stated that the issue had been resolved and the bleeding resolved on its own.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because no sample was returned.A potential root cause for this failure could be "mechanical failure".It was unknown whether the device had met relevant specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "intended use: the catheter is intended for urinary bladder drainage in patients requiring catheterization for management of incontinence, voiding dysfunction, and surgical procedures.Please contact your physician to determine which product options are best for you, paying close attention to product warnings/ precautions and adverse reactions.The catheter becomes slippery when wetted with water, eliminating the need for a separate lubricant.For your added convenience, this catheter is packaged with its own sterile water.Simply release the water from its foil packet and then tip the un-opened catheter package end-to-end.The catheter acts like a magnet to attract the water and activate its slippery coating.Follow these steps for best results.1.Release the water prior to opening the sealed catheter pouch: a.Apply pressure to the foil packet to release the water.B.Ensure all water is released from the foil packet.2.Wet the catheter a.Hold package with printed side up.B.Tip package end-to-end three to six times to wet catheter.This movement is required so that the water transfers back and forth over the catheter to fully wet the hydrophilic coating.3.Use the catheter a.Peel back package to expose funnel end of catheter.B.If desired, use the self-adhesive tape on the package to temporarily attach the pouch to any dry vertical surface.C.Remove catheter and use according to physician¿s instructions.Warning: this is a single use device.Do not reuse.Reuse of a single use device increases the risk of catheter acquired urinary tract infections." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|
| |
|
Search Alerts/Recalls
|