| Model Number |
9001G |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained via: "when were 3 x sutures dislodged from swage (in the package, during removal from package, during handling prior to use on patient or during use on the patient)? please specify during handling.Any fragments left inside patient ? no".Events reported via: 2210968-2023-00038 and 2210968-2023-00040.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown procedure on (b)(6) 2022 and suture was used.During the procedure, 3 x sutures dislodged from swage.No adverse patient consequences were reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 2/3/2023.H6 component code: g07002 no device problem found.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection and functional test evaluation were conducted on the returned devices.Visual analysis of the returned sample determined that it was received fourteen unopened samples that pertain to the product code 9001g.As per the sampling plan, a visual inspection was performed on thirteen samples, the packets were opened, and the swage and attachment area was noted to be as expected.The sutures were dispensed without problems and examined along the strand no anomalies were observed during the evaluation.A functional test was performed using instron equipment and the pull force result was above the minimum requirements.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|