• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MELSUNGEN AG INTROCAN SAFETY® 3; CATHETER,INTRAVASCULAR,THE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MELSUNGEN AG INTROCAN SAFETY® 3; CATHETER,INTRAVASCULAR,THE Back to Search Results
Model Number 4251128-02
Medical Device Problem Code Fail-Safe Problem (2936)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
As reported by the user facility: "the iv was started and had a flash into the base of the iv set.I went to pull the needle out of the catheter and it would not pull out.I was pulling straight back.I put pressure onto the catheter wings and pulled the needle straight back and it still would not release.I pushed down on the catheter more and used more force to pull the needle back and it finally released after a couple of tries.When the needle came out of the catheter, the needle guard was not on the end of the needle.It was three quarters of the way down the needle, toward the base.I believe this device was defective.".
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).A photo was provided for further evaluation.Based on the evaluation results, it was observed that the location of the location of the safety clip is in the middle of the cannula and not covering the cannula.However, without the actual sample to evaluate the exact root cause was unable to be determined.The manufacturer evaluated their production processes and were unable to identify a process that would create this defect.The actual defective device is a valuable tool in investigating the cause of this incident.Review of the batch history records was also performed for the reported lot number and no non-conformances or deviations were noted during the manufacturing process or final inspections.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTROCAN SAFETY® 3
Common Device Name
CATHETER,INTRAVASCULAR,THE
Manufacturer (Section D)
B. BRAUN MELSUNGEN AG
carl-braun strabe 1
melsungen, 34212
GM  34212
Manufacturer (Section G)
B. BRAUN MELSUNGEN AG
carl-braun strabe 1
melsungen, 34212
GM   34212
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key16093797
Report Number9610825-2022-00552
Device Sequence Number15129549
Product Code FOZ
UDI-Device Identifier04046964014017
UDI-Public(01)04046964014017
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K111236
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/01/2026
Device Model Number4251128-02
Device Catalogue Number4251128-02
Device Lot Number21L02G8371
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/20/2022
Initial Report FDA Received Date01/04/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-