| Model Number |
SONATA |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
12/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow-up report.
|
| |
|
Event or Problem Description
|
|
The parents reported that the user had a head trauma with bleeding in the (b)(6) 2022, afterwards he showed poor auditory responses with the device.The user has been re-implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: damage to the active electrode which is consistent with minute device mobility was determined to have led to device failure over time due to fatigue breakages.Also, an impact to the head might have weakened the fixation of the active electrode and could also be a factor for electrode mobility.The investigation results appear to match the problems mentioned in the recipient report.This is a final report.
|
| |
|
Event or Problem Description
|
|
The parents reported that the user had a head trauma with bleeding in the (b)(6) 2022, afterwards he showed poor auditory responses with the device.The user has been re-implanted.
|
| |
|