|
Visual and functional inspections were performed on the returned device.The reported deflation issue could not be confirmed; however, the reported difficulty removing the device was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.It was reported that the nc trek bdc was overinflated 7-8 times to 24 atmospheres (atms).It should be noted that the coronary dilatation catheters (cdc), nc trek rx, global, instructions for use (ifu) states: balloon pressure should not exceed the rated burst pressure (rbp).The rbp for the nc trek rx device is 18 atm.In this case, it is unknown if the ifu violation caused or contributed to the reported complaints.The investigation was unable to determine a conclusive cause for the reported deflation issue and difficulty removing the device from the guiding catheter.Possible factors that may contribute to difficulty deflating the balloon include, but are not limited to, deflation technique, tortuous anatomy, loose connection with the indeflator, contamination in the inflation lumen or damage to the guide wire and/or inflation lumen.Additionally, possible factors that can contribute to difficult to remove/resistance with the guiding catheter post-inflation include, but are not limited to, loose balloon folds, guiding catheter lumen obstructions, kinks, bends or procedural technique.In this case, a conclusive cause for the reported deflation issue and difficulty removing the device from the guiding catheter could not be determined.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|