| Model Number |
IPN915191 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/15/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
|
| |
|
Event or Problem Description
|
|
It was reported that during a surgery, a clip got broken at ligation, another clip was not loaded to the applier properly, and the other clip got broken at the hinge at loading.The user managed to complete the surgery with the applier.The fragment of the clip fell in the patient, but all of them were removed and discarded.No patient injury occurred.The user made an inquiry to us that the clips might have been defective first, but when our sales representative visited the hospital and checked the applier, he found that the jaws were slightly deformed.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The instrument with lot w1-01 was shipped to tfx with invoice (b)(4) of 22/11/21.Original manufacturing lot is 22142129, hr review: upon review of the device history records for the affected lot number, we can confirm that both the correct material and correct components had been used and that the instrument meets the product specifications.All process steps were found to have been properly documented.100% functional test at final inspection found good.The dimensional check did not reveal any deviating dimensions.Clips were picked up and applied without any problems, and the visual inspection showed no evidence of a bent jaw.A bent pull wire was detected, which had no effect on the pick-up and application of clips.The instrument is working properly and is not the cause of the error.Possible causes are user errors, as indicated by the bent pull wire due to strong force, or faulty clips.No further measures will be initiated by (b)(4); this complaint is being rejected by us.
|
| |
|
Event or Problem Description
|
|
It was reported that during a surgery, a clip got broken at ligation, another clip was not loaded to the applier properly, and the other clip got broken at the hinge at loading.The user managed to complete the surgery with the applier.The fragment of the clip fell in the patient, but all of them were removed and discarded.No patient injury occurred.The user made an inquiry to us that the clips might have been defective first, but when our sales representative visited the hospital and checked the applier, he found that the jaws were slightly deformed.
|
| |
|
Search Alerts/Recalls
|