| Model Number |
10005877 |
| Medical Device Problem Code |
Device Difficult to Setup or Prepare (1487)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/16/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Related manufacturer's report: (b)(4).No further information was provided.A supplemental report will be submitted when the manufacturer's investigation is completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a pump exchange on (b)(6) 2022.A new heartmate 3 was implanted used the existing apical cuff.During implant, the original mini cuff was separating from the myocardium while the heartmate 3 (hm3) was secured.The surgeon then used the unlocking tool for the hm3 to remove it from the cuff.Additional pledgeted sutures were placed to secure and achieve hemostasis of the mini apical cuff.Once the cuff was reinforced, the implanting surgeon locked the new hm3 to the cuff, confirming yet again that the yellow lines disappeared and that the hm3 was secured to the apical cuff.The surgeon relayed that the pump was not engaged with the cuff even though the yellow lines were not showing.Some bleeding was noted.After breaking scrub, the surgeon stated that the sintered titanium was slightly visible between the hm3 and center ring of the mini cuff.The pump was unlocked again with the tool and re-secured in the mini cuff, being sure it was fully engaged.The third securement with the mini cuff was successful in achieving hemostasis.The physician proceeded with implant without further issues.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer's investigation conclusion: the reported difficulty engaging the pump with the mini apical cuff could not be confirmed through this evaluation.The reported engagement issue would have prevented the inflow cannula o-ring from sealing against the mini apical cuff hardware, resulting in the reported bleeding.A specific cause for this event could not be conclusively determined.The patient remains ongoing heartmate 3 left ventricular assist system, serial number (b)(6), and will complete a 6-week course of antipseudomonal therapy.The relevant sections of the device history records were reviewed and showed no deviations from manufacturing or quality assurance specifications.Heartmate 3 mini apical cuff kit instructions for use (ifu) (rev.E) is currently available.The section entitled ¿surgical procedures¿ explains how to prepare the mini apical cuff and the apical cuff holder for use.Section ¿directions for use¿ explains how to secure the mini apical cuff to the ventricle.This section also instructs the user that the mini apical cuff should be affixed only by the sew-then-cut method and pledgets should be placed no more than one and a half centimeters from the edge of the felt.Affixing the mini apical cuff using methods other than the prescribed technique can lead to challenges engaging the slide lock mechanism.The heartmate 3 left ventricular assist system (lvas) ifu (rev.C) provides instructions on all surgical procedures, including preparing the ventricular apex site and inserting the pump in the ventricle.The ifu cautions that after the apical cuff has been sewn to the heart, the metal ring on the apical cuff should extend above the felt surface to allow proper engagement and locking with the slide lock of the heartmate 3 left ventricular assist device (lvad).If the slide lock mechanism on the heartmate 3 lvad fails to engage, do not make further attempts to engage until retracting the slide lock mechanism.Evidence of slide lock mechanism failing to engage will be either visual evidence of the yellow ¿wings¿ or a tactile feel of three ridges versus one.The slide lock will not engage the apical cuff unless initially fully retracted.If difficulty persists when engaging the slide lock mechanism, the lvad should be removed from the apical cuff to visualize what might be preventing the connection.The suture knots should not interfere with the connection.If a sealing agent is used on or near the apical cuff, it should not interfere with the slide lock mechanism.During the implant process, a complete backup system (implant kit and external components) must be available on-site and in close proximity for use in an emergency.No further information was provided.The manufacturer is closing the file on this event.
|
| |
|
Event or Problem Description
|
|
It was later reported that the bleeding had resolved.The patient was stable and progressing with plans to complete 6 weeks of antipseudomonal therapy from surgery.Related mfr number for the pump exchange: 2916596-2022-15445.Related mfr number for the new pump: 2916596-2023-00033.
|
| |
|