| Model Number |
0500318N |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
|
| |
|
Event or Problem Description
|
|
A user facility biomedical technician (biomed) reported a dialyzer blood leak that occurred during a patient¿s hemodialysis (hd) treatment.Additional information was obtained through follow-up with the facility.A nurse familiar with the event stated the blood leak occurred approximately two hours and ten minutes into the treatment.The blood leak was reported to be internal.The nurse stated the blood leak was visually observed in the dialysate compartment of the dialyzer.Blood streaks were noted on the outside of the fiber bundle.The machine, a fresenius 2008t machine, alarmed appropriately with a blood leak alarm.B.Braun bloodlines were being utilized for the treatment.A blood test strip was used, and it tested positive for the presence of blood.No visible damage was observed on the dialyzer.After the machine alarmed, the treatment was paused.The patient¿s blood was not returned; their estimated blood loss (ebl) was 250 ml.The nurse confirmed there was no patient injury, no adverse effects were experienced, and no medical intervention was required as a result of the reported event.The patient completed treatment after being resetup with new supplies on a different machine.The complaint device was not available to be returned for manufacturer evaluation as it was reportedly discarded.
|
| |
|
Event or Problem Description
|
|
A user facility biomedical technician (biomed) reported a dialyzer blood leak that occurred during a patient¿s hemodialysis (hd) treatment.Additional information was obtained through follow-up with the facility.A nurse familiar with the event stated the blood leak occurred approximately two hours and ten minutes into the treatment.The blood leak was reported to be internal.The nurse stated the blood leak was visually observed in the dialysate compartment of the dialyzer.Blood streaks were noted on the outside of the fiber bundle.The machine, a fresenius 2008t machine, alarmed appropriately with a blood leak alarm.B.Braun bloodlines were being utilized for the treatment.A blood test strip was used, and it tested positive for the presence of blood.No visible damage was observed on the dialyzer.After the machine alarmed, the treatment was paused.The patient¿s blood was not returned; their estimated blood loss (ebl) was 250 ml.The nurse confirmed there was no patient injury, no adverse effects were experienced, and no medical intervention was required as a result of the reported event.The patient completed treatment after being resetup with new supplies on a different machine.The complaint device was not available to be returned for manufacturer evaluation as it was reportedly discarded.
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: the reported complaint was not confirmed as the complaint device was not returned for manufacturer evaluation.A production records review was performed on the reported lot.An investigation of the device history records (dhr) was conducted by the manufacturer.There was no indication of product nonacceptance, deviation, non-conformance, rework, labeling or process control failure during the manufacturing process which could be associated with the reported event.The lot met all release criteria.A definitive conclusion regarding the complaint incident cannot be reached without physical examination of the actual device.Therefore, the complaint is not confirmed.Continuous improvement is of the utmost importance to fresenius medical care as we strive to provide dialysis products of the highest quality to our patients.Reports of leaking product are investigated both individually as complaints, as well as via the nc/capa program, in order to assess and improve our products and processes.Capas for vision systems and blood leak reduction are recent examples of leak related investigations directed at an overall reduction in dialyzer leaks.
|
| |
|