MAUDE Adverse Event Report: OGDEN MANUFACTURING PLANT OPTIFLUX 180NR DIALYZER FINISHED ASSMBLY; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OGDEN MANUFACTURING PLANT OPTIFLUX 180NR DIALYZER FINISHED ASSMBLY; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Model Number 0500318N
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
 
Event or Problem Description
A user facility biomedical technician (biomed) reported a dialyzer blood leak that occurred during a patient¿s hemodialysis (hd) treatment.Additional information was obtained through follow-up with the facility.A nurse familiar with the event stated the blood leak occurred approximately two hours and ten minutes into the treatment.The blood leak was reported to be internal.The nurse stated the blood leak was visually observed in the dialysate compartment of the dialyzer.Blood streaks were noted on the outside of the fiber bundle.The machine, a fresenius 2008t machine, alarmed appropriately with a blood leak alarm.B.Braun bloodlines were being utilized for the treatment.A blood test strip was used, and it tested positive for the presence of blood.No visible damage was observed on the dialyzer.After the machine alarmed, the treatment was paused.The patient¿s blood was not returned; their estimated blood loss (ebl) was 250 ml.The nurse confirmed there was no patient injury, no adverse effects were experienced, and no medical intervention was required as a result of the reported event.The patient completed treatment after being resetup with new supplies on a different machine.The complaint device was not available to be returned for manufacturer evaluation as it was reportedly discarded.
 
Event or Problem Description
A user facility biomedical technician (biomed) reported a dialyzer blood leak that occurred during a patient¿s hemodialysis (hd) treatment.Additional information was obtained through follow-up with the facility.A nurse familiar with the event stated the blood leak occurred approximately two hours and ten minutes into the treatment.The blood leak was reported to be internal.The nurse stated the blood leak was visually observed in the dialysate compartment of the dialyzer.Blood streaks were noted on the outside of the fiber bundle.The machine, a fresenius 2008t machine, alarmed appropriately with a blood leak alarm.B.Braun bloodlines were being utilized for the treatment.A blood test strip was used, and it tested positive for the presence of blood.No visible damage was observed on the dialyzer.After the machine alarmed, the treatment was paused.The patient¿s blood was not returned; their estimated blood loss (ebl) was 250 ml.The nurse confirmed there was no patient injury, no adverse effects were experienced, and no medical intervention was required as a result of the reported event.The patient completed treatment after being resetup with new supplies on a different machine.The complaint device was not available to be returned for manufacturer evaluation as it was reportedly discarded.
 
Additional Manufacturer Narrative
Plant investigation: the reported complaint was not confirmed as the complaint device was not returned for manufacturer evaluation.A production records review was performed on the reported lot.An investigation of the device history records (dhr) was conducted by the manufacturer.There was no indication of product nonacceptance, deviation, non-conformance, rework, labeling or process control failure during the manufacturing process which could be associated with the reported event.The lot met all release criteria.A definitive conclusion regarding the complaint incident cannot be reached without physical examination of the actual device.Therefore, the complaint is not confirmed.Continuous improvement is of the utmost importance to fresenius medical care as we strive to provide dialysis products of the highest quality to our patients.Reports of leaking product are investigated both individually as complaints, as well as via the nc/capa program, in order to assess and improve our products and processes.Capas for vision systems and blood leak reduction are recent examples of leak related investigations directed at an overall reduction in dialyzer leaks.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLUX 180NR DIALYZER FINISHED ASSMBLY
Common Device Name
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
OGDEN MANUFACTURING PLANT
director, site quality
475 west 13th street
ogden UT 84404
Manufacturer (Section G)
OGDEN MANUFACTURING PLANT
director, site quality
475 west 13th street
ogden UT 84404
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key16097810
Report Number0001713747-2023-00004
Device Sequence Number6669242
Product Code KDI
UDI-Device Identifier00840861100125
UDI-Public00840861100125
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K002277
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 01/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2025
Device Model Number0500318N
Device Catalogue Number0500318N
Device Lot Number22KU04015
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/27/2022
Supplement Date Received by Manufacturer01/10/2023
Initial Report FDA Received Date01/04/2023
Supplement Report FDA Received Date01/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FRESENIUS 2008T MACHINE; FRESENIUS 2008T MACHINE
Patient Age61 YR
Patient SexFemale
Patient Weight131 KG
-
-