• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION BURETROL SOLUTION SETS; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION BURETROL SOLUTION SETS; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number ARC7503P
Device Problems Fluid/Blood Leak (1250); Fitting Problem (2183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/06/2022
Event Type  malfunction  
Event Description
It was reported that the spike of a buretrol solution set did not fit with a non-baxter bag which caused the unspecified solution to leak.The issue was identified during patient use.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Initial reporter address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
The actual device was not available; however, a photograph of the sample was provided for evaluation.A visual inspection was performed to the photograph which did not identify any abnormalities that could have contributed to the reported condition.The reported condition is not clearly observed via the provided photographs and due to the nature of the returned sample no additional testing could be performed.Therefore, the reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BURETROL SOLUTION SETS
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CALI
calle 36 no. 2c-22,
apartado aero 2446
cali
CO  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key16098049
MDR Text Key308476262
Report Number1416980-2022-07228
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00085412479484
UDI-Public(01)00085412479484
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberARC7503P
Device Lot NumberSE22HJ3
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/09/2022
Initial Date FDA Received01/04/2023
Supplement Dates Manufacturer Received02/05/2023
Supplement Dates FDA Received02/08/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/15/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
LR FRESENIUS BAG
-
-