ABBOTT VASCULAR ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM Back to Search Results
Model Number 1012534-80
Medical Device Problem Codes Difficult or Delayed Activation (2577); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the procedure was to treat a lesion in the right superficial femoral artery with mild calcification and mild tortuosity.The 6.0x80mm absolute pro self-expanding stent system (sess) was advanced to the target lesion with a.035 supracore guidewire; however during deployment the thumbwheel ratchet was advanced with resistance.Although deployment was difficult the stent was implanted successfully.There was no adverse patient effect and no clinically significant delay reported in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.As there was no damage noted to the device during the inspection prior to use, the investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that interaction with the mildly calcified and mildly torturous anatomy and/or other devices resulted in restricting the shaft lumens from moving freely; thus, preventing the thumbwheel from rotating, resulting in the reported physical resistance / sticking and the reported difficult or delayed activation.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM
Common Device Name
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key16103446
Report Number2024168-2023-00162
Device Sequence Number12676066
Product Code NIP
UDI-Device Identifier08717648176005
UDI-Public08717648176005
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P110028
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2025
Device Model Number1012534-80
Device Catalogue Number1012534-80
Device Lot Number2041261
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/14/2022
Initial Report FDA Received Date01/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/12/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
.035 SUPRACORE GUIDEWIRE
Patient SexUnknown
-
-