• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. DERMABOND MINI HV 0.36ML VIAL - 12EA; ADHESIVE, TOPICAL SKIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. DERMABOND MINI HV 0.36ML VIAL - 12EA; ADHESIVE, TOPICAL SKIN Back to Search Results
Model Number DHVM12
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Component code: (b)(4).Additional information has been requested and the following was received.Attempts to obtain the device have been made.If further details are received at a later date a supplemental will be sent.What was done to treat the shard penetration? was medical or surgical intervention required? was there any special diagnostic test performed? what is the status of the surgeon injury today? was it difficult to break the ampoule (was extra force needed to break the ampoule)? was the ampoule crushed repeatedly? can you identify the lot number of the product that was used? is the product involved in the event or a representative sample (product from the same lot number) available for evaluation? if yes, what is the device return status? this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported a patient underwent a cabg procedure on (b)(6) 2022 and topical skin adhesive was used.During use,when mixing the product the ampule broke and a shard of glass pierced the assistant's glove.No injury was sustained to the user.No patient consequences reported.
 
Additional Manufacturer Narrative
Product complaint # (b)(4) additional information: d4, d9, h4.H3 evaluation: the device was returned for analysis.Upon visual inspection, it was observed a crushed ampoule and dried formulation inside the bulb.The filter is purple in color due to contact with the formulation.The sample reveals a piercing through which a glass shard protruded in the applicator bulb.Also, a piercing in the butyrate tube is observed.These piercings coincided in position.Visual inspection shows that all the components of the applicator are present and appropriately assembled.A manufacturing record evaluation was performed for the finished device batch, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device history review
=
> n/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DERMABOND MINI HV 0.36ML VIAL - 12EA
Common Device Name
ADHESIVE, TOPICAL SKIN
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN LORENZO PR
road 183, km. 8.3
san lorenzo 00754
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key16109335
Report Number2210968-2023-00125
Device Sequence Number7654273
Product Code MPN
UDI-Device Identifier10705031203525
UDI-Public10705031203525
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K152096
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/12/2024
Device Model NumberDHVM12
Device Catalogue NumberDHVM12
Device Lot NumberSKBBTZ
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/20/2022
Supplement Date Received by Manufacturer01/06/2023
Initial Report FDA Received Date01/05/2023
Supplement Report FDA Received Date01/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/12/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-