| Model Number |
68LP-480 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
11/09/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated. if any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time. reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side deflation.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side deflation.Device remains implanted.
|
| |
|
Search Alerts/Recalls
|