• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH INTELLIVUE MULTI MEASUREMENT SERVER SLCP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH INTELLIVUE MULTI MEASUREMENT SERVER SLCP Back to Search Results
Model Number 866462
Medical Device Problem Code Failure to Sense (1559)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/29/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Reporting address line 1: (b)(6).Reporting institution phone #: (b)(6).Reporter phone #: (b)(6).
 
Event or Problem Description
The customer reported a neonatal disposable blood oxygen probe measurement error after one hour of measurement; the probe can't measure oximetry (spo2) after one hour.The device was in clinical use at the time the issue was discovered.There was no adverse event or patient harm reported.On 22dec2022, the philips distributor investigating the case found the spo2 was lost in isolation, and there were no error messages or alarms from other parameters that alerted the clinical staff to the problem.The distributor confirmed the spo2 sensor was faulty and replaced the spo2 sensor.The issue was resolved.The distributor advised the customer to order a replacement sensor.
 
Event or Problem Description
The customer reported a neonatal disposable blood oxygen probe measurement error after one hour of measurement; the probe can't measure oximetry (spo2) after one hour.The device was in clinical use at the time the issue was discovered.There was no adverse event or patient harm reported.On 22dec2022, the philips distributor investigating the case found the spo2 was lost in isolation, and there were no error messages or alarms from other parameters that alerted the clinical staff to the problem.The distributor confirmed the spo2 sensor was faulty and replaced the spo2 sensor.The issue was resolved.The distributor advised the customer to order a replacement sensor.Additional information was received on 10feb2023 indicating that there was an inoperative error message which had alerted the user to this issue.This is no longer consider a reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIVUE MULTI MEASUREMENT SERVER SLCP
Common Device Name
INTELLIVUE MULTI MEASUREMENT SERVER SLCP
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key16111805
Report Number9610816-2023-00006
Device Sequence Number14866703
Product Code MHX
UDI-Device Identifier00884838055094
UDI-Public00884838055094
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number866462
Device Catalogue Number866462
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date01/06/2023
Supplement Report FDA Received Date02/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/23/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-