• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM (VIPERWIRE ADVANCE); DIAMONDBACK PERIPHERAL ORBITAL ATHERECTOMY SYSTEM (VIPERWIRE ADVANCE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM (VIPERWIRE ADVANCE); DIAMONDBACK PERIPHERAL ORBITAL ATHERECTOMY SYSTEM (VIPERWIRE ADVANCE) Back to Search Results
Model Number VPR-GW-14
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 12/08/2022
Type of Reportable Event Serious Injury
Event or Problem Description
A diamondback 360 peripheral exchangeable orbital atherectomy device (oad) was used to treat a 5mm, calcified lesion in the mid popliteal.There was a previously placed stent in the distal side of the popliteal.The viperwire advance guide wire was placed in the distal side of the stent.The oad was used to begin treatment of the mid popliteal.After two or three treatments, the oad was advanced too close to the viperwire and fractured the spring tip, as the viperwire was unable to be advanced more distally.The spring tip was snared.A new viperwire and non-csi devices were used to complete the procedure without further complication.
 
Additional Manufacturer Narrative
The viperwire advance guide wire was returned with a fracture within the spring tip.Scanning electronic micrscopic imaging showed evidence of fatigue.The root cause of the event was determined to be due to use not consistent with the ifu.The peripheral ifu states: make sure there is a minimum of 10cm between the guide wire spring tip and the distal end of the shaft.The material inspection report for this guide wire lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.Csi id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM (VIPERWIRE ADVANCE)
Common Device Name
DIAMONDBACK PERIPHERAL ORBITAL ATHERECTOMY SYSTEM (VIPERWIRE ADVANCE)
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer (Section G)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer Contact
aaron stevens
1225 old hwy 8 nw
st. paul, MN 55112
MDR Report Key16113825
Report Number3004742232-2023-00003
Device Sequence Number6839114
Product Code MCW
UDI-Device Identifier10852528005077
UDI-Public(01)10852528005077(17)240731(10)449185-1
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K190634
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2024
Device Model NumberVPR-GW-14
Device Catalogue Number72023-01
Device Lot Number449185-1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/08/2022
Initial Report FDA Received Date01/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/25/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age81 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-