| Model Number |
87-3435 |
| Medical Device Problem Codes |
Fracture (1260); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
12/20/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Patient was revised due to the femoral head dislocated, once head was removed, surgeon noticed constrained liner had been cracked.Replaced constrained liner and implanted a 28+12 srom head.Original implant date was unknown.It was unknown if any loosening occurred.There was no reported surgical delay.Doi: unknown.Dor: (b)(6) 2022.Affected side: left hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.
|
| |
|
Search Alerts/Recalls
|