| Model Number |
66-2509 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device is coming apart.No injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Protocol number is unknown.No product information has been provided to date.A product sample was received and is awaiting evaluation and investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
Other text: one unit was returned for investigation.Upon physical inspection, it was found that the complained issue could be duplicated.The sample was being pulled apart at less than 10 lbs and the specification sheet states that the parts should withstand 10 lbs.The problem source is that the assembler not properly heat staking parts together.Dhr review was done, no issues related to the original complaint were found.
|
| |
|