• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEFIBTECH, L.L.C. LIFELINE; CARDIAC COMPRESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEFIBTECH, L.L.C. LIFELINE; CARDIAC COMPRESSOR Back to Search Results
Model Number RMU-1000
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code Unspecified Heart Problem (4454)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An inspection of the returned device identified that the the root cause was attributed to loosening of the rmu-1000 motor's bottom screw.
 
Event or Problem Description
An international distributor reported that when a customer was using the arm with the patient, the device stopped working.No patient details were provided.They reported that after the event they checked the device's report and found the following warnings: "warn - motor should be home" and "warn - motor didn't move in time".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFELINE
Common Device Name
CARDIAC COMPRESSOR
Manufacturer (Section D)
DEFIBTECH, L.L.C.
741 boston post rd.
suite 201
guilford CT 06437
Manufacturer (Section G)
DEFIBTECH, L.L.C.
741 boston post road
suite 201
guilford CT 06437
Manufacturer Contact
ian white
741 boston post road
suite 201
guilford, CT 06437
2034536654
MDR Report Key16117472
Report Number3003521780-2023-00001
Device Sequence Number5667081
Product Code DRM
UDI-Device Identifier10815098020628
UDI-Public10815098020628
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K141809
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Consumer,Health Professional,Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberRMU-1000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/01/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/01/2022
Initial Report FDA Received Date01/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/26/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-