Model Number UHI-4
Device Problem Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/06/2022
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned to olympus for evaluation.The investigation is ongoing.If additional information becomes available, this report will be supplemented accordingly.
 
Event Description
An olympus representative reported to olympus, on behalf of the customer, the high flow insufflation unit front panel was not working and it couldn¿t be operated during maintenance.There was no patient involvement associated with this event.
 
Manufacturer Narrative
Updated: d8, d9, e4, h3, h4, h6, h10.This report is being submitted to provide additional information based on the device evaluation results.The olympus service center found that the reported ¿the front panel is not functioning and cannot be operated¿ issue was confirmed, and was due to corrosion of the flat cable.Two additional reportable malfunctions were also identified during the evaluation 1.After changing the flat cable, the display was found to not be fully visible and needs to be replaced.2.A tube was found damaged and needs to be replaced.In addition, the evaluation found the following non-reportable device defects: corrosion on foot switch connector, minor corrosion on the rear panel and corrosion on the main board.The non-reported device defects identified in the device evaluation have no potential to cause or contribute to death or serious injury if the malfunctions were to recur.The investigation is ongoing, therefore, the root cause of the reported issues cannot be determined at this time, however, if additional information becomes available this report will be supplemented accordingly.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 2 years since the subject device was manufactured.Based on the results of the investigation, the cause as to why the flat cable was corroded could not be identified, and the cause of the faulty front panel unit could not be identified.In regards to the damaged tube, the root cause of the phenomenon could not be identified.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16117641
MDR Text Key308583708
Report Number3002808148-2023-00186
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 04/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/06/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/10/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/19/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-