• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BELMONT MEDICAL TECHNOLOGIES CRITICOOL; THERMAL REGULATING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BELMONT MEDICAL TECHNOLOGIES CRITICOOL; THERMAL REGULATING SYSTEM Back to Search Results
Catalog Number 508-03521
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/15/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The internal complaint file#: (b)(4) has been logged for this incident for traceability.We received the curewrap involved in this incident back for investigation on (b)(6) 2022.Belmont received a medwatch report#: (b)(4) on december 7th, 2022.The initial report from the user detailed the incident that occurred 11/15.They noticed linens soaked with water and there was water on the floor.The user detailed no impact to patient since they quickly swapped out the wrap.The medwatch report#: (b)(4) detailed the product problem that curewrap infant cooling blanket acquired multiple holes in it and soaked baby and bed linen.Rn had to replace the cooling blanket during the rewarming process.The set temperature was increased to 35.3 at 22:00 and core temperature reading 35.0 so within normal variation for set temperature.The patient involved in the incident was unharmed.The investigation to date has indicated that the issue with wrap is attributable to a change in the component manufacturing process parameters by our supplier's sub-supplier.Belmont has issued scar (supplier corrective action request) to the supplier in relation to this issue and is working with the material to issue a process change and additional testing to avoid any further issues in future.As of november 2022, the sub-supplier confirmed that they reverted to the original process parameters and the manufacturing process is back in control.Wraps that belmont received starting november 2022 were manufactured per the correct manufacturing parameters.All wrap products manufactured during the identified period of the manufacturing parameter change and within the control of belmont, have been quarantined.6 lots have been fully consumed.A review of the associated risk management has been conducted and it has been confirmed that the risk involved in the incident is low and the severity is minor.The operator's manual informs the user, "if moisture or leaks are discovered in the connecting hose and/or wrap, turn off the criticool device, disconnect the power cable from its power source, and correct the problem before proceeding." the manual also provides a troubleshooting guide for water leaks, as well as the following warning statement: "water may drip from the inlet tubes of the wraps.Be sure that no electrical device or outlet is located under the criticool's water inlet or wrap tubes.When disconnecting wraps from the criticool confirm that the clamps are tight, to prevent water leaking from the wrap."all curewraps are 100% leak tested by the manufacturer prior to release for shipment.The user facility reported that there were no clinical consequences for the patient.Belmont is currently investigating the returned curewrap.Without results of the device investigation, no final conclusions can be drawn.A follow-up report will be submitted once the investigation is complete and additional information becomes available.
 
Event or Problem Description
The user detailed the incident occurred 11/15.The user noticed linens soaked with water and there was water on the floor.They detailed no impact to patient since they quickly swapped out the wrap.The patient involved in the incident was unharmed.
 
Additional Manufacturer Narrative
Upon receipt of the referenced curewrap, having s/n#: (b)(6) and lot# m220761 & lot# m222131, belmont service engineers confirmed the failures by using leak testing, performed functional testing for maximum use time of the wraps, and performed burst testing. the root cause was determined to be a change in the manufacturing process of the water barrier at a supplier of the material.If a leak occurs, the patient may get wet, however thermoregulation will continue.In the event thermoregulation cannot continue due to the leak, the device will alarm to alert the user.The wrap can be replaced to continue thermoregulation.The change led to unfavorable properties of the material and a potential of the material having microscopic pinholes that were not detectable during the manufacturing and testing process.The lots involved in the reported complaints were determined to have been manufactured with this compromised material.The failures were associated with lots that were using material produced by the sub supplier, with modified process parameters.Belmont issued a scar# (b)(4) (supplier corrective action request) was issued to the supplier (creative foam) on (b)(6) 2022 to the supplier to resolve this issue.The following actions were taken by the supplier: 1.Investigated internal manufacturing processes utilizing samples provided by customers after failure and unused products from lots that reported a leak.The investigation focused on the wrap sealing process.The investigation ruled out heat sealing of the wrap as the root cause.2.The water barrier material was sent out for investigation to an independent laboratory and the report indicated a difference between material supplied prior to 2021 and the material used in the listed lots.3.Communication with sub supplier of the water barrier material concluded that the sub supplier introduced a very slight change to the manufacturing process which affected the properties of the material, leading to thinner areas that had the potential to leak.4.The supplier implemented a no change agreement with the sub supplier to ensure that no changes without prior approval would be introduced in the future and the sub supplier returned the process to their original specifications.All wraps are 100% leak tested by pressure decay test at the end of the manufacturing process and prior to packaging.An additional testing performing wet (destructive) test was introduced at the supplier as part of the corrective action.Belmont concluded that lots of wraps manufactured with water barrier material produced with the modified process parameters could include wraps with a weak material area that potentially could result in leaks during use.Belmont visited the user facility on (b)(6) 2022 to gather additional information and ensure that the replaced wraps are working as intended without any issues.Upon discussion with the clinical staff, they confirmed that the newly sent wraps are working without any problems we will continue to monitor this type of incident and take further corrective and preventive actions if required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRITICOOL
Common Device Name
THERMAL REGULATING SYSTEM
Manufacturer (Section D)
BELMONT MEDICAL TECHNOLOGIES
780 boston road
billerica MA 01821
Manufacturer (Section G)
BELMONT MEDICAL TECHNOLOGIES
780 boston road
billerica MA 01821
Manufacturer Contact
urja jani
780 boston road
billerica, MA 01821
MDR Report Key16118596
Report Number1219702-2023-00001
Device Sequence Number6802179
Product Code DWJ
UDI-Device Identifier17290012127196
UDI-Public17290012127196
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K083662
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 03/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number508-03521
Device Lot NumberM220761
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/16/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/07/2022
Supplement Date Received by Manufacturer12/07/2022
Initial Report FDA Received Date01/06/2023
Supplement Report FDA Received Date03/10/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-