|
Reported event: an event regarding crack/fracture involving a restoration modular stem was reported.The event was confirmed through a material analysis of the device and review of medical records provided.Method & results: product evaluation and results: visual inspection: visual inspection was performed as part of the material analysis report (mar).This inspection indicated 'visual and stereo-microscope examination image shows the bowed conical distal stem with fracture location highlighted by the arrow.On visual examination, the fracture was observed near the distal tip of the stem.The cone body remained attached to the distal stem and the femoral head remained attached to the cone body.Stereo microscope imaging of the fracture surface associated with the proximal end of the stem.Based on macroscopic surface features observed, including beach marks, the approximate location of origin (o) and direction of propagation (p) were determined.The macroscopic surface features observed on the fracture surface indicate the component fractured due to fatigue.' dimensional inspection: not performed as the dimensional aspects are not in question.Functional inspection: not performed as the functional aspects are not in question.Material analysis: a material analysis has been performed.The report concluded: 'electron microscopy examination shows a scanning electron microscope image of the approximate fracture origin location, associated with the proximal end of the bowed conical distal stem.The fracture propagated in fatigue as indicated by the typical fatigue striations observed at the origin and found throughout the fracture surface, as shown in representative secondary electron image.Energy dispersive spectroscopy (eds) was performed on the base material to characterise its material composition.Elemental constituents included ti, al and v which are consistent with titanium 6ai-4v eli alloy [1].Discussion & conclusion review of the bowed conical distal stem, catalogue # 6276-7-216, lot code caxt70w confirmed a fracture at the distal end of the stem.Characterisation using stereo microscopy and scanning electron microscopy confirmed fracture of the stem near the distal end due to fatigue.No manufacturing or material related defects were observed on the device surfaces examined.' medical records received and evaluation: a review of the provided medical records by a clinical consultant indicated: as described above, this patient has undergone four different hip arthroplasty procedures including three revisions since 2003.The second revision was with a restoration modular implant and the third revision was with a long stem cemented exeter implant.Conclusion of assessment: this patient has undergone multiple surgeries for loosening of a total hip femoral stem and subsequently with fracture of two different types of femoral stems.I can confirm that these events occurred since i was able to review x-rays documenting these procedures.However i did not see the preoperative x-ray for the third revision.I can see that the distal portion of the stem was left in situ.I cannot determine the root cause of this event with certainty.The causes of femoral stem loosening and fracture are multifactorial including surgical technique factors, patient bmi and activity level, which in this case is a significant contributing factor, and implant factors.The explanted implant should be returned to stryker for engineers to perform an analysis.A complete analysis is not possible however because the retained portion of the stem will not be able to be returned.Product history review: a review of the device manufacturing records indicate that all devices accepted into final stock were free from discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the mar and clinician review of provided medical records confirmed the stem fracture.The mar report concluded: 'electron microscopy examination shows a scanning electron microscope image of the approximate fracture origin location, associated with the proximal end of the bowed conical distal stem.The fracture propagated in fatigue as indicated by the typical fatigue striations observed at the origin and found throughout the fracture surface, as shown in representative secondary electron image.Energy dispersive spectroscopy (eds) was performed on the base material to characterise its material composition.Elemental constituents included ti, al and v which are consistent with titanium 6ai-4v eli alloy [1].' the clinician review concluded : ;this patient has undergone multiple surgeries for loosening of a total hip femoral stem and subsequently with fracture of two different types of femoral stems.I can confirm that these events occurred since i was able to review x-rays documenting these procedures.However i did not see the preoperative x-ray for the third revision.I can see that the distal portion of the stem was left in situ.I cannot determine the root cause of this event with certainty.The causes of femoral stem loosening and fracture are multifactorial including surgical technique factors, patient bmi and activity level, which in this case is a significant contributing factor, and implant factors.The explanted implant should be returned to stryker for engineers to perform an analysis.A complete analysis is not possible however because the retained portion of the stem will not be able to be returned.; no further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|