| Model Number |
COOLCUT 90,ASP ABLATOR |
| Medical Device Problem Code |
Smoking (1585)
|
| Health Effect - Clinical Codes |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the head of the device sparked and smoked and charred.There was no harm for patient, operator or third party reported.This was noticed before the surgery.No further information received.Update (b)(6) 2023.It was confirmed that the device was used with a generator from another manufacturer.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
One unpackaged ar-9803a-90 arthrex coolcut®, aspirating ablator, 3 mm, 90 serial/batch number: (b)(6) was received for investigation.Functional testing was not performed because the device connector switch is incompatible with the testing equipment.Visual inspection did not identify issues with the device.However, by the complaint event description, the equipment was used: "with a generator from another manufacturer (erbe)" according to dfu-0088-2, rev.0 monopolar ablation probe.
|
| |
|