| Model Number |
M00510860 |
| Medical Device Problem Codes |
Difficult to Advance (2920); Material Deformation (2976); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The returned trapezoid rx basket was analyzed, and a visual evaluation noted that the side car rx was pushed back approximately 5.5 mm which is out of specification.There is also insufficient evidence to claim that the device was difficult to advance into the endoscope and that no further failures were discovered on the device.The reported event was not confirmed.Based on all available information, there is insufficient evidence to conclude that the device was difficult to advance into the endoscope, and no further failures were discovered on the device; thus, the reported complaint issues " device difficult to advance & side car - rx torn material " are not confirmed and will be documented as no problem detected.The side car being pushed back may have been caused due to an excess of force was applied during the manipulation.Perhaps the technique used could have contributed to the reported event.Therefore, the most probable root cause is adverse event related to procedure.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a trapezoid rx basket was used in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2022.During the procedure, the guidewire lumen tip was flattened, which appeared to be the cause of the guidewire's inability to insert smoothly.Another trapezoid basket was used to complete the procedure.There were no patient complications reported as a result of this event.Investigation results revealed the side car rx was pushed back; therefore, this is now an mdr reportable event.
|
| |
|
Search Alerts/Recalls
|