• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH 2.7MM TI VA LCKNG SCR SLF-TPNG WITH T8 STARDRIVE RECESS 20MM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH 2.7MM TI VA LCKNG SCR SLF-TPNG WITH T8 STARDRIVE RECESS 20MM; SCREW, FIXATION, BONE Back to Search Results
Catalog Number 04.211.020TS
Medical Device Problem Codes Break (1069); Migration (4003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Part #04.211.020ts, lot #8l95358.Manufacturing site: (b)(4).Release to warehouse date: 06 jan 2022.Expiration date: 01 jan 2027.Supplier: (b)(4).Non-sterile part # 04.211.020, non-sterile lot # 546p876, manufacturing site: (b)(4).Release to warehouse date: 07 dec 2021.Supplier: (b)(4).Additional procode: hrs.Complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial , a follow-up will be filed as appropriate.
 
Event or Problem Description
Device report from japan reports an event as follows: it was reported that on (b)(6) 2022, the patient underwent orif for clavicular diaphysis with the products in question.Two months later, on (b)(6) 2022 screw breakage and backout was discovered.As of (b)(6) 2022, the outside is intact, and the patient is being monitored for bone fusion.No further information is available.This report is for a 2.7mm ti va lckng scr slf-tpng with t8 stardrive recess 20mm.This is report 1 of 2 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.7MM TI VA LCKNG SCR SLF-TPNG WITH T8 STARDRIVE RECESS 20MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES SELZACH
bohackerweg 5
selzach 2545
SZ   2545
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 
3035526892
MDR Report Key16121285
Report Number8030965-2023-00303
Device Sequence Number8557829
Product Code HWC
UDI-Device Identifier10886982119025
UDI-Public(01)(01)10886982119025
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K100776
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number04.211.020TS
Device Lot Number8L95358
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/14/2022
Supplement Date Received by Manufacturer01/19/2023
Initial Report FDA Received Date01/09/2023
Supplement Report FDA Received Date01/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/06/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CORTSCR ø2.7 SELF-TAP L18 TAN; VA-LCP CLAVIC PL 2.7 SHAFT CS1 R TAN
Outcome Attributed to Adverse Event Other;
Patient Age42 YR
Patient SexMale
-
-