| Model Number |
IPN915322 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).N/a.Other remarks: n/a.Corrected data: n/a.
|
| |
|
Event or Problem Description
|
|
It is reported that after the catheter was inserted, the balloon was found to be ruptured.A second catheter was inserted at the same insertion site.No medical intervention necessary.The patient status is reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).No serial number was reported.The serial number on the returned sample is (b)(6).The lot number (18f22f0051) recorded on the complaint report matches the lot number on the returned original packaging box and for the returned sample.Returned for investigation was a 40cc 7.5fr ultraflex intra-aortic balloon catheter (iabc) with an original packaging box that matches the serial number on the returned sample.The sample was returned in a cardboard box and was loosely packed within the original packaging box.Returned with the sample was supplied data-scope inflation driveline tubing.Upon return, the one-way valve was tethered to the short driveline tubing.The bladder was fully unwrapped.The iabc outer lumen was noted damaged with flattened surface at approximately 29.2cm to 31cm from the iabc distal tip.Dried blood was noted on the exterior surfaces of the returned iabc.No blood was noted within the helium pathway.No other visual damage or abnormalities were noted to the returned sample.The bladder thickness was measured at six points with measurements ranging from 0.0057in-0.0065in.The one-way valve was tested and passed.A vacuum was pulled on the one-way valve, and it held for at least 1 minute and then 30 seconds five separate times.The iabc central lumen was aspirated and flushed using a 60cc lab-inventory syringe.Some blood exited.The iabc was leak tested.No leaks were detected.Full inflation was achieved.The device passed the leak test.The iabc was connected to an iabp with a lab inventory 40cc inflation driveline tubing.The iabc was inserted into the "t" tube and 100mmhg backpressure was applied.The iabc was pumped for a minimum of 15 minutes.There were no alarms triggered.The bladder inflated and deflated completely with each beat.The balloon pressure waveform (bpw) was normal.A lab inventory 0.025in guidewire was back loaded through the iabc distal tip.No resistance was noted; the guidewire was able to advance through the central lumen.Some blood noted on the guidewire.The guidewire was front loaded through the iabc luer.No resistance was noted; the guidewire was able to advance through the central lumen.Some blood noted on the guidewire.A device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.The risk is acceptable.The returned device passed visual and functional testing.The reported complaint of iab leak suspected is not confirmed.During the visual inspection of the returned sample, no blood was noted within the helium pathway.The returned intra-aortic balloon catheter (iabc) was functionally tested with no leaks or alarms noted.The returned device passed visual and functional test specifications.Based on a review of the device history record (dhr), the product met specification upon release.The root cause of the complaint is undetermined.No further action required at this time.Other remarks: n/a.Corrected data: n/a.
|
| |
|
Event or Problem Description
|
|
It is reported that after the catheter was inserted, the balloon was found to be ruptured.A second catheter was inserted at the same insertion site.No medical intervention necessary.The patient status is reported as "fine".
|
| |
|
Search Alerts/Recalls
|