MAUDE Adverse Event Report: ST. JUDE MEDICAL ADVISOR¿ VL, SENSOR ENABLED¿ CIRCULAR MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL ADVISOR¿ VL, SENSOR ENABLED¿ CIRCULAR MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number D-AVSE-DF20-V1525
Medical Device Problem Codes Entrapment of Device (1212); Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/20/2022
Type of Reportable Event Malfunction
Event or Problem Description
During the paroxysmal atrial fibrillation procedure using the advisor¿ vl, sensor enabled¿ circular mapping catheter, it was noted that it became entangled with the chordae tendineae.The catheter was inserted into the left atrium, and the catheter inserted into the left ventricle.When the geometry of the left atrium was being created, it was noted the catheter was entangled with the chordae tendineae, and it could not come out.The catheter was rotated, but it was still hard to remove, so the sheath was removed.No damage was seen on the catheter, but it was replaced and the procedure was completed with no adverse consequences to the patient.
 
Additional Manufacturer Narrative
One 22 pole, bi-directional, curve d-f, sensor enabled, advisor vl circular mapping catheter was received for evaluation.The catheter loop was stretched and bent.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported removal difficulty and entanglement remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVISOR¿ VL, SENSOR ENABLED¿ CIRCULAR MAPPING CATHETER
Common Device Name
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key16122891
Report Number3008452825-2022-00742
Device Sequence Number20443377
Product Code DRF
UDI-Device Identifier05415067027283
UDI-Public05415067027283
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K192037
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2025
Device Model NumberD-AVSE-DF20-V1525
Device Catalogue NumberD-AVSE-DF20-V1525
Device Lot Number8399252
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/20/2022
Supplement Date Received by Manufacturer02/10/2023
Initial Report FDA Received Date01/09/2023
Supplement Report FDA Received Date03/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/24/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-