| Model Number |
407450 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
During an atrial fibrillation ablation procedure, an air leak was noted.The sheath was inserted into the right atrium of the patient and prior to inserting the catheters and a transseptal needle, it was noted that air was aspirated into a syringe that connected to the extension port of the sheath.Several aspirations were attempted but the air was still aspirated into the syringe.The sheath was replaced and the procedure was completed with no adverse consequences to the patient.No additional femoral vein puncture was required for replacing the sheath.
|
| |
|