| Model Number |
IPN048873 |
| Medical Device Problem Code |
Deflation Problem (1149)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
12/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Reported issue: the ed physician intubated a patient and had difficulty with the tube.Respiratory checked the balloon prior to intubation and it inflated properly.After introducing the tube, it would not hold air.This happened on 2 different et tubes, 3rd held air.The tube is a 7.5mm hudson rci tube, lot 73d220280 with a 04/10/2027 expiration.This same situation happened with two other 8.0 tube et tubes.The reported defect was detected during use on patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).The device history review for the et tube, cf,7.5 lot # 73d2200280 investigation did not show issues related to the complaint.The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
|
| |
|
Search Alerts/Recalls
|