• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG CASPAR DISTRACTION PIN 14MM; SPINE SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG CASPAR DISTRACTION PIN 14MM; SPINE SURGERY Back to Search Results
Model Number FF904SB
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to aesculap aag that a caspar distraction pin 14mm (item # ff904sb) was used during an anterior cervical discectomy and fusion (acdf) procedure performed on (b)(6) 2022.According to the complainant, during the procedure, the distraction pin broke off into the cervical vertebrae and could not be removed.The procedure was completed with another aesculap distraction pin.The top tip of the complaint device was returned to the manufacturer for evaluation.A fragment remained in situ.The adverse event is filed under aag reference (b)(4).
 
Additional Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
Update: the malfunction is filed under aag reference (b)(4).
 
Additional Manufacturer Narrative
Investigation results: visual investigation: we received a distraction pin (ff904sb) with broken off tip.We made a visual inspection of the fracture surface.Here we found the typical signs of a multi axial stress condition of bending and torsion.The yellow lines marking the bending and torsion lines, the red hatched area marks the residual forced fractured area.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.The review of risk assessment revealed that the overall risk level (severity 2(5) x probability 2(5) of occurrence) according to din en iso 14971 is still acceptable.Explanation and rationale: the lot 52768652 contains 4000 pins.This is the only complaint out of this lot.The dhr shows no signs of deviation to the specifications during production.The error is typically for mechanical overstress of a multi axial stress condition of bending and torsion.A material failure or a manufacturing error can be ruled out.Conclusion and measures / preventive measures: based upon the investigation results the root cause is most probably mechanical overstress.There is no indication for a material-, manufacturing- or design-related failure.Based upon the investigations results a capa is not necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CASPAR DISTRACTION PIN 14MM
Common Device Name
SPINE SURGERY
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
christian von der grün
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key16125125
Report Number9610612-2023-00006
Device Sequence Number6548069
Product Code LXH
UDI-Device Identifier04038653364305
UDI-Public4038653364305
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 03/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFF904SB
Device Catalogue NumberFF904SB
Device Lot Number52768652
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/05/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/13/2022
Supplement Date Received by Manufacturer03/03/2023
Initial Report FDA Received Date01/09/2023
Supplement Report FDA Received Date03/29/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/28/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-