This report is being submitted to relay additional information and device evaluation.The following sections are being reported: h6: investigation type codes were added: 3331, 4109, 4110 and 4111.H6: investigation findings code was added: 180.H6: investigation conclusions codes were added: 4307.One (1) twist ha 3.25mmx10mm (2130) was returned for investigation.Visual evaluation of the as returned product identified signs of use.Based on the evaluation, device malfunction did occur.However, there is no existing nonconformance/capa/hhe/d/ie/product holds against the reported product that did or could cause or contribute to the reported event.Monthly post market trending review identified no actionable trends or corrective actions for the reported event or device.Zimvie quality management system (qms) has controls in place to ensure the distribution of conforming product.Therefore, based on the available information, the product was within specifications and likely conforming when it left zimvie.Dhr review was completed for the subject lot number (2020120704).It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review was performed for the reported lot number (2020120704) for similar events and no other complaint was identified.Functional testing to recreate the reported event; verified the implant and mount could not disengage.A definitive root cause could not be identified.However, based on the investigation, ifu and risk file review, the most likely causes determined from the investigation are clinician error, implant not placed properly, excessive insertion torque.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
|