• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHT MEDICAL TECHNOLOGY INC INBONE TALAR DOME SZ 3 SULCUS; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHT MEDICAL TECHNOLOGY INC INBONE TALAR DOME SZ 3 SULCUS; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 220220903
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 12/14/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.Device was discarded.
 
Event or Problem Description
It was reported that the patient presented with an unknown infection.The patient underwent a total ankle system removal procedure.The infection is still active.
 
Event or Problem Description
It was reported that the patient presented with an unknown infection.The patient underwent a total ankle system removal procedure.The infection is still active.
 
Additional Manufacturer Narrative
The reported event could be confirmed, based on available medical records and health care professionals.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.The instructions for use instructs user that: ¿intraoperative and early postoperative complications can include: 8) deep wound infection (early or late) which may necessitate removal of the prosthesis.On rare occasions, arthrodesis of the involved joint or amputation of the limb may be required.¿ formal medical opinion was sought from an experienced independent medical expert and he opined below: ¿the tar components forming the device have been explanted because of an infection, the components were replaced by a hand-made cement-spacer.Since the device was not retrieved, and it is not present on the revision ct, their state cannot be assessed.¿ ¿without further information, we may assume that the cause of the infection is patient-related and/or procedure related, not-device related, until proven.¿ more detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If device is returned or any further information is provided, the investigation report will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INBONE TALAR DOME SZ 3 SULCUS
Common Device Name
PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key16126361
Report Number3010667733-2023-00008
Device Sequence Number16947036
Product Code HSN
UDI-Device Identifier00840420124227
UDI-Public00840420124227
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K100886
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number220220903
Device Catalogue Number220220903
Device Lot Number1706459
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/14/2022
Supplement Date Received by Manufacturer06/16/2023
Initial Report FDA Received Date01/09/2023
Supplement Report FDA Received Date07/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age43 YR
Patient SexMale
Patient Weight82 KG
-
-