• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) AGILIS HISPRO¿ STEERABLE CATHETER WITH ELECTRODES; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) AGILIS HISPRO¿ STEERABLE CATHETER WITH ELECTRODES; CATHETER, PERCUTANEOUS Back to Search Results
Model Number DS3H010-38
Medical Device Problem Code Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Code Cardiac Perforation (2513)
Date of Event 12/20/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Related manufacturing reference number: 2017865-2023-00728.It was reported that the patient presented for an implant procedure.During the procedure, it was noted that the right ventricular lead perforated the left ventricle.The physician alleged that the perforation was caused by the high pull back strength of the catheter.The lead was repositioned successfully.No further intervention was performed.The patient will continue to be monitored.There were no patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AGILIS HISPRO¿ STEERABLE CATHETER WITH ELECTRODES
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16127189
Report Number2017865-2023-00730
Device Sequence Number11572831
Product Code DQY
UDI-Device Identifier05415067032287
UDI-Public05415067032287
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K200721
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2022
Device Model NumberDS3H010-38
Device Catalogue NumberDS3H010-38
Device Lot Number8309077
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/20/2022
Initial Report FDA Received Date01/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/25/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-