| Model Number |
101/870/070CZ |
| Medical Device Problem Codes |
Deflation Problem (1149); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during the use of the product, the pilot balloon got detached from the inflation line, and the cuff was found deflated.No patient injury reported.The reported lot number couldn't be verified.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: one unit was returned for investigation.Upon physical inspection, it was found that the complained issue could be duplicated.The inflation line was detached from the pilot balloon and the complaint is confirmed.A dhr (device history review) was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.No manufacturing or service issues were identified as causes of the customer's reported problem during the review of the service and repair records.
|
| |
|