Model Number 101/870/070CZ
Medical Device Problem Codes Deflation Problem (1149); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/01/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during the use of the product, the pilot balloon got detached from the inflation line, and the cuff was found deflated.No patient injury reported.The reported lot number couldn't be verified.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Additional Manufacturer Narrative
Other, other text: one unit was returned for investigation.Upon physical inspection, it was found that the complained issue could be duplicated.The inflation line was detached from the pilot balloon and the complaint is confirmed.A dhr (device history review) was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.No manufacturing or service issues were identified as causes of the customer's reported problem during the review of the service and repair records.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX TUBES BLUSELECT
Common Device Name
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key16129857
Report Number3012307300-2023-00299
Device Sequence Number11572859
Product Code BTO
UDI-Device Identifier15019517076202
UDI-Public15019517076202
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K173384
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number101/870/070CZ
Device Catalogue Number101/870/070CZ
Device Lot Number4065614
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/16/2022
Supplement Date Received by Manufacturer01/27/2023
Initial Report FDA Received Date01/10/2023
Supplement Report FDA Received Date02/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
Patient SexUnknown
-
-