Model Number 101/870/075CZ
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/01/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that when the customer was performing a suction for the top of the cuff, the suction line connector got detached.No patient injury reported.
 
Additional Manufacturer Narrative
No product information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Additional Manufacturer Narrative
Conclusion codes: updated device evaluation: one used, decontaminated device was received for investigation.Visual inspection noted a piece of suction line tubing was still inside the connector (bonded with the connector).The suction line tubing was separated from connector near the connector entrance.The device was sent to a second investigation site for further root cause analysis.At the second investigation site only the connector was received, the connector was detached from the tube and a piece of tube was identified inside the connector, confirming the complaint.Based on the analysis conducted in the sample and photos provided of the sample, the occurrence of this failure condition could be caused by the incorrect use of the product, since a portion of the suction line tubing remains inside the connector, indicating that the union was correct and due to an external force, the tubing was pulled out of the connector.A review of the device history record (dhr) could not be completed as no lot number was provided.No corrective actions are required since the main cause of the detachment problem is improper use of the product.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX TUBES BLUSELECT
Common Device Name
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key16129879
Report Number3012307300-2023-00300
Device Sequence Number11576448
Product Code BTO
UDI-Device Identifier15019517076219
UDI-Public15019517076219
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K173384
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number101/870/075CZ
Device Catalogue Number101/870/075CZ
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/16/2022
Supplement Date Received by Manufacturer05/11/2023
Initial Report FDA Received Date01/10/2023
Supplement Report FDA Received Date05/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
Patient SexUnknown
-
-