| Model Number |
101/870/075CZ |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that when the customer was performing a suction for the top of the cuff, the suction line connector got detached.No patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
No product information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion codes: updated device evaluation: one used, decontaminated device was received for investigation.Visual inspection noted a piece of suction line tubing was still inside the connector (bonded with the connector).The suction line tubing was separated from connector near the connector entrance.The device was sent to a second investigation site for further root cause analysis.At the second investigation site only the connector was received, the connector was detached from the tube and a piece of tube was identified inside the connector, confirming the complaint.Based on the analysis conducted in the sample and photos provided of the sample, the occurrence of this failure condition could be caused by the incorrect use of the product, since a portion of the suction line tubing remains inside the connector, indicating that the union was correct and due to an external force, the tubing was pulled out of the connector.A review of the device history record (dhr) could not be completed as no lot number was provided.No corrective actions are required since the main cause of the detachment problem is improper use of the product.
|
| |
|
Search Alerts/Recalls
|