| Model Number |
3152 |
| Medical Device Problem Codes |
Loss of or Failure to Bond (1068); Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/20/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The scissor was returned with the shrink tubing removed from the device.The shrink tubing does not appear to be damaged.It appears that the shrink tubing was not adequately shrunk to remain tightly affixed to the scissor hub.An investigation and capa: (b)(4) have been initiated to determine the root cause and corrective action.
|
| |
|
Event or Problem Description
|
|
During the surgery, heat shrink came off and fell into the abdominal cavity.It was removed from the abdominal cavity and there was no health problem for the patient.
|
| |
|