• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number 3544475
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Capsular Contracture (1761); Breast Implant Associated Anaplastic Large Cell Lymphoma (BIA ALCL) (4549)
Date of Event 02/25/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported by that a 52-year-old caucasian female who underwent primary breast augmentation with a 450cc mentor memorygel breast implant and a 475cc mentor memorygel breast implant experienced an anaplastic large cell lymphoma diagnosis post procedure.The patient presented with a left groin mass.A biopsy was performed on (b)(6) 2022 and the final diagnosis was anaplastic large cell lymphoma.As a result, explant was performed on (b)(6) 2022.See 1645337-2023-00337 for contralateral prosthesis report.
 
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: anaplastic large cell lymphoma.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
Manufacturer¿s reference number: (b)(4).On january 10, 2023 mentor became aware that the incorrect event date was erroneously submitted with the initial report.The correct event date is february 25, 2022.
 
Additional Manufacturer Narrative
Additional information was received on january 20, 2023.The patient weight was obtained and populated in a4.In addition to the issues reported initially, the patient also experienced capsule contracture (baker grade unknown).The capsular contracture was confirmed through a surgical pathology report conducted on the breast capsule explanted on (b)(6) 2022.The results of the pathology report have been populated in b6.Additional clinical review of this case was performed on (b)(6) 2023. per pathology, tissue sections reveal a high-grade malignant neoplasm composed of pleomorphic large cells.Initial immunostains confirm a cd30-positive lymphoproliferative disorder.Additional stains confirm the coexpression of alk in the absence of all other pertinent markers.Thus, while the differential diagnosis included anaplastic large cell lymphoma, classical hodgkin lymphoma, and peripheral t-cell lymphoma, nos, the latter two have been excluded by additional immunostaining results.The overall combined morphologic features (large pleomorphic/anaplastic cells) and immunophenotype (cd30 positive, alk-positive) are diagnostic of anaplastic large cell lymphoma (alcl). however, the alcl is not breast implant-associated.Reason for device explant and/or reoperation: capsular contracture.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key16134880
Report Number1645337-2023-00338
Device Sequence Number6717624
Product Code FTR
UDI-Device Identifier00081317003946
UDI-Public00081317003946
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/02/2020
Device Model Number3544475
Device Catalogue Number3544475
Device Lot Number6993389
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/10/2023
Supplement Date Received by Manufacturer01/10/2023
01/20/2023
Initial Report FDA Received Date01/10/2023
Supplement Report FDA Received Date01/10/2023
02/02/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/03/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Required Intervention;
Patient Age52 YR
Patient SexFemale
Patient Weight68 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-