BIOMET MICROFIXATION 45MM RIGHT NARROW MANDIBULAR COMPONENT; TMJ SYSTEM Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Abscess (1690); Bacterial Infection (1735); Pain (1994); Loss of Range of Motion (2032); Swelling/ Edema (4577); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/17/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported the patient underwent a bilateral tmj procedure.Subsequently, the patient was revised on the right side due to unknown reasons.
 
Additional Manufacturer Narrative
Zimmer biomet complaint (b)(4).Medical products item# 24-6562 , lot# 976690b, tmj right fossa component, small.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001032347-2022-00403.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: contributing factors of the event included bilateral advanced tmj arthritis, right mastoiditis, revision tmj, and chronic otitis media.Patient experienced pain, lock jaw, ringing in the ears, swelling, edema, erythema, difficulty opening mouth, and possible cellulitis.Right fossa and mandible prosthesis exposed with purulence and were explanted.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was further reported in the patient's medical records that an initial bilateral tmj procedure took place on an unknown date over 30 years ago with unknown components.A first revision occurred due to degenerative arthritis, ankylosis, trismus, and shortening of mastication muscle.Approximately two months later, the patient developed swelling, pain, edema, crepitus, and decreased range of motion.The patient underwent a two-stage revision.During the 1st stage revision, the presence of gross purulence from abscess that surrounded the condyle and fossa components were noted.The right condyle and fossa were explanted and two drains were placed.A picc line placed 2 days later for iv antibiotics for 6 weeks, followed by oral antibiotics for bacterial infection.The 2nd stage revision included implantation of replacements of condyle and fossa, screws for maxillomandibular fixation, and autogenous fat graft to the tmj region without complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
45MM RIGHT NARROW MANDIBULAR COMPONENT
Common Device Name
TMJ SYSTEM
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer (Section G)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key16135184
Report Number0001032347-2022-00404
Device Sequence Number6659914
Product Code LZD
UDI-Device Identifier00841036081386
UDI-Public(01)00841036081386(17)250720(10)986920B
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P020016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/20/2025
Device Model NumberN/A
Device Catalogue Number01-6545
Device Lot Number986920B
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/14/2022
Supplement Date Received by Manufacturer01/10/2023
Initial Report FDA Received Date01/10/2023
Supplement Report FDA Received Date02/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/20/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age46 YR
Patient SexFemale
Patient Weight73 KG
-
-