Model Number 1456Q/86
Device Problem Separation Failure (2547)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/22/2022
Event Type  malfunction  
Manufacturer Narrative
Further information was requested, but not received.
 
Event Description
It was reported that the patient presented for an implant procedure.During the procedure, it was noted that the stylet was failing to be removed from the left ventricle (lv) lead.The left ventricle lead was not used.The physician continued to use the second left ventricle lead and completed the procedure.The patient was discharged post-procedure.
 
Manufacturer Narrative
The reported event of ¿stylet stuck¿ was confirmed.A complete lead was returned in two pieces.Visual inspection found the polytetrafluoroethylene (ptfe) coating of the stylet bunched up/clogged distal to the connector pin.X-ray examination also found bunching of the inner coil at this location.The cause of the reported event was due to the bunching of the stripped stylet, ptfe coating and inner coil.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET LEAD SMALL-S, 86 CM
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16135356
MDR Text Key307077333
Report Number2017865-2023-00870
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734510158
UDI-Public05414734510158
Combination Product (y/n)Y
Reporter Country CodeIN
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 03/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/10/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1456Q/86
Device Catalogue Number1456Q/86
Device Lot NumberA000127440
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/10/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/25/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-