| Model Number |
MI1210 SYNCHRONY ST |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected and channels with abnormal impedances have been observed.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected and channels with abnormal impedances have been observed.Re-implantation is considered, however a date for surgery has not been scheduled yet.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: based on the received information, a damage to the active electrode due to excessive mechanical stress appears very likely.However, to determine an exact root cause device investigation would be necessary.Re-implantation is considered but no date has been scheduled yet.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: based on the received information, a damage to the active electrode due to excessive mechanical stress appears very likely.However, to determine an exact root cause device investigation would be necessary.The concerned device was explanted but has not been received for investigation yet.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected and channels with abnormal impedances have been observed.The user was re-implanted.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected and channels with abnormal impedances have been observed.The user was re-implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: overload fractures in the active electrode which are consistent with an external mechanical impact were determined to be the root cause of device failure.The problems given in the recipient report appear to match the damage found.This is a final report.
|
| |
|