MAUDE Adverse Event Report: MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number MI1210 SYNCHRONY ST
Medical Device Problem Code Impedance Problem (2950)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event or Problem Description
The user's hearing performance with the device is affected and channels with abnormal impedances have been observed.
 
Event or Problem Description
The user's hearing performance with the device is affected and channels with abnormal impedances have been observed.Re-implantation is considered, however a date for surgery has not been scheduled yet.
 
Additional Manufacturer Narrative
Additional information: based on the received information, a damage to the active electrode due to excessive mechanical stress appears very likely.However, to determine an exact root cause device investigation would be necessary.Re-implantation is considered but no date has been scheduled yet.
 
Additional Manufacturer Narrative
Additional information: based on the received information, a damage to the active electrode due to excessive mechanical stress appears very likely.However, to determine an exact root cause device investigation would be necessary.The concerned device was explanted but has not been received for investigation yet.
 
Event or Problem Description
The user's hearing performance with the device is affected and channels with abnormal impedances have been observed.The user was re-implanted.
 
Event or Problem Description
The user's hearing performance with the device is affected and channels with abnormal impedances have been observed.The user was re-implanted.
 
Additional Manufacturer Narrative
Conclusion: overload fractures in the active electrode which are consistent with an external mechanical impact were determined to be the root cause of device failure.The problems given in the recipient report appear to match the damage found.This is a final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM
Common Device Name
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key16136381
Report Number9710014-2023-00019
Device Sequence Number20472842
Product Code MCM
UDI-Device Identifier09008737331832
UDI-Public(01)09008737331832
Combination Product (Y/N)N
Initial Reporter CountryTU
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 10/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/17/2021
Device Model NumberMI1210 SYNCHRONY ST
Device Catalogue Number33334
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/09/2023
Supplement Date Received by Manufacturer01/09/2023
01/09/2023
01/09/2023
Initial Report FDA Received Date01/11/2023
Supplement Report FDA Received Date03/22/2023
09/25/2023
10/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/17/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age7 YR
Patient SexFemale
-
-