• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TPRLC 133 T1 PPS SO 10X140MM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TPRLC 133 T1 PPS SO 10X140MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Concomitant medical products: unknown broach.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation has been completed, a follow-up mdr will submitted.
 
Event or Problem Description
It was reported that the surgeon had difficulty with the stem not seating to the exact broach size.The broaches went down to where they were supposed to; however, the implant sat between 1-4 mm proud.The surgeon would then drop down 2-3 sizes and rebroach.After that, the implant would sit in place.There were no patient complications.It was reported that no further information is available.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.H6: component code: mechanical (g04) - stem no product was returned, or pictures provided; visual and dimensional evaluation could not be performed.Medical records were not provided.A review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC 133 T1 PPS SO 10X140MM
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key16136866
Report Number0001825034-2023-00011
Device Sequence Number17774509
Product Code LZO
UDI-Device Identifier00880304489684
UDI-Public(01)00880304489684(17)320305(10)7197874
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K200196
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number51-103100
Device Lot Number7197874
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/30/2022
Supplement Date Received by Manufacturer01/12/2023
Initial Report FDA Received Date01/11/2023
Supplement Report FDA Received Date01/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-