NUVASIVE, INC. NUVASIVE NEXT GENERATION NVM5 SYSTEM; NEUROSURGICAL NERVE LOCATOR Back to Search Results
Model Number 50120104
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/22/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Following a case in which interoperative neuromonitoring was used, a needle electrode broke off inside the patient's skin during needle removal.The surgeon surgically excised the broken needle to remove it form the patient, which also delayed wakeup by approximately 1 hour.
 
Additional Manufacturer Narrative
Three (3) radiographs were provided and used to confirm the reported event.Each radiograph showed the fractured portion of the of electrode retained in the patient's skin.Additionally, two (2) photographs of the complaint device were also provided, which showed the electrode wire and connection hub with approximately 1 to 2 mm of the metal electrode tip remaining on the device and the tip of the electrode absent.Based on the information obtained, the root cause of the reported event is unknown, but is likely the result of repeated bending or flexion of the electrode needle during placement and removal combined with handling and electrode removal techniques, which caused a breakage of the solid metal component of the electrode needle.No additional investigation can be completed at this time.Labelling review: "warnings, cautions and precautions.¿proper handling, insertion and placement of electrodes is critical for monitoring." ".Improperly placed or bent needles increase the risk of needle breaking off in the patient." ".Do not attempt to straighten bent needles because this may cause stress and weaken the needle causing it to break off in the patient." "¿do not implant the instruments: complications to the patient may include, but are not limited to.Breakage of the device, which could make necessary removal difficult or sometimes impossible, with possible consequences of late infection and migration.Breakage could cause injury to the patient.Injury that may result from instrument breakage, slippage, misuse, or mishandling.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE NEXT GENERATION NVM5 SYSTEM
Common Device Name
NEUROSURGICAL NERVE LOCATOR
Manufacturer (Section G)
NUVASIVE, INC.
7475 lusk blvd.
san diego CA 92121
Manufacturer Contact
geoffrey gannon
7475 lusk blvd.
san diego, CA 92121
MDR Report Key16137611
Report Number2031966-2023-00005
Device Sequence Number6610264
Product Code PDQ
UDI-Device Identifier00887517939869
UDI-Public887517939869
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K162313
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/02/2023
Device Model Number50120104
Device Lot Number1110767
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/22/2022
Initial Report FDA Received Date01/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/14/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NUVASIVE PULSE III SYSTEM
Outcome Attributed to Adverse Event Required Intervention;
Patient SexPrefer Not To Disclose
-
-