| Model Number |
N/A |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/20/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during an initial right knee surgery, the tibial impactor pad broke during trial impacting.Surgery was completed using different pads.There was no surgical delay.The surgical technique was utilized.No foreign bodies were retained.No known impact or consequence to the patient.Attempts have been made and all available information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Proposed component code: mechanical (g04)- impactor.Visual examination of the returned product identified signs of repeated use (nicked and gouged) and component has fractured.All pieces were not returned.The device history record was reviewed and no discrepancies relevant to the reported event were found.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Search Alerts/Recalls
|