(b)(6).A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: visual analysis of the returned device identified: weight to the spec, broken shell, part of the shell is missing, crease fold.A microscopic analysis was performed which identify a striated edge broken.Based on the device analysis the final assessment is: a striated edge broken on radius side assessed as surgical damage.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been/will be requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade unknown.
|