| Model Number |
DL-344-10-K |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Pending investigation of returned lead.
|
| |
|
Event or Problem Description
|
|
The patient's right depth lead displayed abnormal signals, saturation, and flat line activity during review of patient ecog data.The abnormal activity appears to have begun around (b)(6) 2022.These activities are indicative of a right depth lead break.The lead was explanted and replaced on (b)(6) 2022.During explant the surgeon reported evidence of lead damage near the burr hole cover.
|
| |
|