As received, a complete lead was returned in one piece with helix found retracted and clogged with blood/tissue.After cleaning, the helix could be extended and retracted by applying torque directly to the connector pin.The full helix extension length was measured to be within specification.A stiffness test was performed, and the results were within specification.Electrical testing did not find any indication of conductor fractures or internal shorts.Visual and x-ray inspections of the lead did not find any anomalies except for procedural damage.
|