• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TIBIAL ARTICULAR SURFACE PROVISIONAL RIGHT SIZES G, H +0 MM THICKNESS BOTTOM; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TIBIAL ARTICULAR SURFACE PROVISIONAL RIGHT SIZES G, H +0 MM THICKNESS BOTTOM; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a tasp fractured during surgery.Attempts to obtain additional information have been made; however, no more information is available.
 
Additional Manufacturer Narrative
(b)(4).The product has been returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up is being submitted to relay additional information.Visual examination of the returned product identified signs of repeated use (nicked or gouged) and is fractured on medial side of post.Medical records were not provided.Device history record (dhr) was reviewed for deviations and/ or anomalies with no deviations / anomalies identified.A definitive root cause cannot be determined.If any further information is which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.H6: component code: proposed component (annex g) code is: mechanical (g04)- provisional bottom.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL ARTICULAR SURFACE PROVISIONAL RIGHT SIZES G, H +0 MM THICKNESS BOTTOM
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key16139645
Report Number0001822565-2023-00108
Device Sequence Number17803506
Product Code JWH
UDI-Device Identifier00889024558168
UDI-Public(01)00889024558168(10)64493525
Combination Product (Y/N)N
Initial Reporter StateMT
Initial Reporter CountryUS
PMA/510(K) Number
K181947
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42589200700
Device Lot Number64493525
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/06/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/19/2022
Supplement Date Received by Manufacturer01/25/2023
Initial Report FDA Received Date01/11/2023
Supplement Report FDA Received Date01/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/12/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexMale
-
-