• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON CARIBE LTD. BD MACRO-VUE¿ RPR CARD TEST; ANTIGENS, NONTREPONEMAL, ALL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON CARIBE LTD. BD MACRO-VUE¿ RPR CARD TEST; ANTIGENS, NONTREPONEMAL, ALL Back to Search Results
Model Number 275005
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It was reported that bd macro-vue¿ rpr card test large particles have been affecting interpretation of results.No injuries were reported.The following information was provided by the initial reporter: the particles are too large, which affects the interpretation.
 
Event or Problem Description
It was reported that bd macro-vue¿ rpr card test large particles have been affecting interpretation of results.No injuries were reported.The following information was provided by the initial reporter: the particles are too large, which affects the interpretation.
 
Additional Manufacturer Narrative
H.6 investigation summary: material 275005, batch 2094445.Satisfactory results when retention samples were tested for performance using three (3) sera panel numbers of each reactivity.Self sample results: control lot (2153970) customer batch number (2094445).Serum no.13: control and retention , showed reactive results at 1:1 to 1:4 and reactive minimal at 1:8.Serum no's; 18: control and retention sample, showed reactive results at 1:1 and reactive minimal at 1:2.Serum no¿s.19: control and retention sample, showed reactive results at 1:1- 1:2 and reactive minimal at 1:4.Serum no¿s.26, 30 - 38: control and retention sample showed reactive minimal at 1:1.Serum no's.7, 8 and 9: all lots showed negative results.Visual inspection was performed to retention samples and no discrepancy were found, no agglutination or particle too course were observed.Normal size of the particle were observed.Batch record review do not show any discrepancy, all testing were within specifications for product release.Photo received from customer do not show agglutination or course particle.Complaint unconfirmed based of retention test results, batch history record and photo evaluation.A cross functional team continually monitors all product complaints for trends and determines if any additional actions are necessary beyond the current investigation process.Special attention should be provided to: antigen storage and use.Sample and room temperature during the test.Type of specimen used: collection and preparation.According of the ¿manual of test for syphilis¿ edition 9.0 the following ¿sources of error¿ had been identified for increase of reactivity or false reactivity for the rpr test.If the time of rotation is too long, test reactivity may be increased.If the card is excessively rotated and tilted (to and fro motions) by hand after removal from the rotator, a false reactive result may occur.If the serum is unevenly spread in the circle, the antigen and the antibody may not mix properly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MACRO-VUE¿ RPR CARD TEST
Common Device Name
ANTIGENS, NONTREPONEMAL, ALL
Manufacturer (Section D)
BECTON DICKINSON CARIBE LTD.
vicks drive
lot no. 6
cayey PR
Manufacturer (Section G)
BECTON DICKINSON CARIBE LTD.
vicks drive
lot no. 6
cayey PR
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key16140545
Report Number3008352382-2023-00002
Device Sequence Number6816957
Product Code GMQ
UDI-Device Identifier00382902750055
UDI-Public00382902750055
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date04/30/2024
Device Model Number275005
Device Catalogue Number275005
Device Lot Number2094445
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/26/2022
Supplement Date Received by Manufacturer01/31/2023
Initial Report FDA Received Date01/11/2023
Supplement Report FDA Received Date02/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/10/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-