Event or Problem Description
It was reported that after centrifugation with bd microtainer® sst¿ there was poor separator movement.There was no report of patient impact.The following information was provided by the initial reporter: it was reported by the customer after centrifuging samples they are not getting any serum.
 
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that after centrifugation with bd microtainer® sst¿ there was poor separator movement.There was no report of patient impact.The following information was provided by the initial reporter: it was reported by the customer after centrifuging samples they are not getting any serum.
 
Additional Manufacturer Narrative
H6.Bd had not received samples or photos for evaluation.Therefore, retention samples of the incident lot were selected from bd inventory for evaluation and upon completion, no issues relating to poor barrier separation were observed.Bd was unable to duplicate or confirm the customer¿s indicated failure mode, poor barrier separation, because the defect was not evident in the testing of the complaint lot samples.Visual evaluations of both retention and control samples demonstrated clinically acceptable performance for all visual observations evaluated.Laboratory evaluation of samples indicated that these tubes performed as expected.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for poor barrier separation.Bd was not able to identify a root cause for the indicated failure mode.Factors that may contribute to poor barrier separation were evaluated through a clinical study to verify the design of the device met it¿s intended use.The result of the study showed that the device performed as expected and we were unable to determine any external contributor to this reported issue.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MICROTAINER® SST¿
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BD CARIBE LTD.
road 31
k.m. 24.3
juncos
Manufacturer (Section G)
BD CARIBE LTD.
road 31
k.m. 24.3
juncos
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key16141392
Report Number2618282-2023-00002
Device Sequence Number6659768
Product Code JKA
UDI-Device Identifier50382903659673
UDI-Public(01)50382903659673
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K991702
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date07/31/2023
Device Model Number365967
Device Catalogue Number365967
Device Lot Number2124289
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/19/2022
Supplement Date Received by Manufacturer03/31/2023
Initial Report FDA Received Date01/11/2023
Supplement Report FDA Received Date04/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/04/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-