| Model Number |
W5-7-3-MVI-3 |
| Medical Device Problem Codes |
Dent in Material (2526); Separation Failure (2547)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/21/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was implanted in the patient and was not returned to the manufacturer for evaluation.Medical imaging was not provided.The alleged product issue could not be confirmed.
|
| |
|
Event or Problem Description
|
|
It was reported that a web device was attempted to be implanted in an aneurysm and would not detach after multiple detachment attempts with three different detachment controllers.The device was detached by invaginating the device in the aneurysm.The proximal recess of the aneurysm was coiled.The patient was unharmed and was discharged same day without complication.
|
| |
|