| Model Number |
ATG80146 |
| Medical Device Problem Codes |
Material Rupture (1546); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/16/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 11/2025).Device pending return.
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure, the balloon allegedly ruptured.It was further reported that the balloon was allegedly broke off into the patient.Reportedly, the broken balloon was retrieved from the patient.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause for the reported balloon rupture and balloon detachment could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: b5, d4 (expiry date: 11/2025),.
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure, the pta balloon allegedly ruptured.It was further reported that the balloon allegedly broke off into the patient.Reportedly, the broken balloon was retrieved from the patient using an equipment.There was no reported patient injury.
|
| |
|